- Up to 128 measurement channels
- High linearity and precision
- Extreme operating temperature -20 to +60 °C (-40 °C option available on request)
- RS-485, RS-232 and Ethernet (TCP/IP)
- Various analog and digital output
- LCD screen display (In Option)
The WellSens Gen II surface controller unit is a robust and reliable multi-channel signal conditioner. It is
compatible with Opsens’ harsh environment interferometric fiber optic pressure and
temperature sensor, the OPP-W. It can operate over a wide temperature range from -40
to +60 °C. Its small foot print, low power consumption, ease of use and its competitive
cost make it the solution of choice for oil fields or industrial process applications.
| Number of channels |
Up to 128 channels |
| Compatibility |
OPP-W, OTP-W, and OEP fiber optic sensors |
| Full scale |
30 000 nm (path length difference) |
| Resolution |
± 0.003 % of F.S. (no averaging) |
| Precision |
± 0.01 % of F.S. @ ± 3.3 sigma limit (99.9 % confidence level) |
| Sampling rate |
Up to 20 Hz |
| Output interface |
RS-485, RS-232 and Ethernet (TCP/IP) |
| Communication protocol |
Modbus and SCPI |
| I/O and control option |
Various analog and digital outputs (4-20 mA, ± 10 V, etc.) |
| Data storage |
Up to 8 Gb (in option) |
| Display |
LCD screen display (in option) |
| Input power |
24 VDC or 85-264 VAC |
| Consumption |
10 W + 1 W x number of channels (Ex.: 16-CH unit = 26 W) |
| Storage temperature |
- 40 °C to 70 °C |
| Operating temperature |
- 20 °C to 60 °C (-40°C option available on request) |
| Humidity |
95 % non condensing |