Signal conditioner
   
AccuSens
page 1
 



Download datasheet




HIGH ACCURACY SIGNAL CONDITIONER FOR MEASURING PHYSIOLOGICAL† TEMPERATURE


Use with Opsens’ OTP-M fiber optic physiological* temperature sensor

 


•  Highest accuracy and resolution in the industry

•  OEM and multi-channel version available

•  Outstanding repeatability

•  20 Hz sampling rate

•  ± 5V and RS-232 output interfaces

 


The AccuSens is a signal conditioner specifically designed for measuring physiological† temperature. The AccuSens is designed to perform in stringent medical environments by offering the highest temperature accuracy and resolution in the industry.

At the heart of the AccuSens is the Opsens’ White Light Polarization Interferometry (WLPI) technology (patent pending) which provides a mean for making accurate and absolute measurements of the path length difference of any type of interferometric fiber optic sensors, whose difference varies according to the measurand of interest.

The AccuSens comes with a standard ± 5V output and a RS-232 communication port for real-time data acquisition. The AccuSens can be controlled directly using the front-panel keypad or remotely using the standard RS-232 interface. Single and multi-channel OEM versions are available.

 
 
 


•  Physiological temperature measurements

•  RF, microwave ad electro surgery environments

•  MRI system (R&D and OEM applications)

•  Sterilization applications


Signal conditioner
   
AccuSens
page 2
 

Number of channels

One (multi-channel OEM version available)

Compatibility

Opsens’ OTP-M fiber optic temperature sensor

Accuracy*

± 0.15 °C

Resolution*

0.01 °C

Sampling rate

20 Hz standard

Output interface

± 5V and RS-232 standard

Input power and consumption

9 to 24 VDC - 1.8W (wall-transformer adapter included)

Dimensions
(without rubber boot protection):
OEM version single channel:


45 mm (H) x 105 mm (W) x 165 mm (L)
30 mm (H) x 90 mm (W) x 155 mm (L)

Storage temperature

-40 °C to 70 °C

Operating temperature

0 °C to 45 °C

Humidity

95 % non condensing

Light source life span

40 000 hours MTBF

 

 

* Specifications include the effect of both the signal conditioner errors and the OTP-M sensor errors.
† The AccuSens product is designed to be integrated into customer host system but is not approved by the FDA or by any applicable regulatory bodies. The compulsory approvals needed for the medical or clinical use of the AccuSens product are the customer’s responsibility.

All specifications are subject to change without prior notification